User's Manual

5.Operation Instructions For Image Acquisition
User Manual of NDT1013LA 59
4. After the XWIN is end, then the detector will start the acquisition flow.
5. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.3.2. Freesync Mode
IDLE
XWIN1
(fixed)
X-Ray
imaging
XWIN1
(fixed)
Offset
imaging
X-Ray Info
RAW Preview
RAW X-Ray image
If HW-offset is disabled
RAW Offset(Or corrected)
image
1. For Freesync mode, there is no low power state.
2. When the detector is Idle, user can start the exposure flow any time.
3. When the X-Ray starts, the detector will sense the X-Ray automaticlly, the XWIN is configured
by parameter “Set Delay Time” on “Detector” page “Parameter” tab, the unit is “ms”, user
needs to make sure that the XWIN is larger than the X-Ray time.
4. After the XWIN is end, then the detector will start the acquisition flow.
5. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.4. After use
1. Disconnect the software
2. Power off
3. Keep it clean
4. Store under specified conditions
5.5. Correction and Calibration Template Generation