User's Manual
5.Operation Instructions For Image Acquisition
User Manual of NDT1013LA 57
HVG
X-Tube
FPD
Work
Station
5.2.2. Work Flow(PrepAcq)
Select HWPostOffset、HWGain、HWDefect. If user need the raw image, please de-select all these
correction options.
Also, the software correction is supported.
IDLE Prepare
XWIN1
(fixed)
X-Ray
imaging
Cmd_PrepAcq Ack_Prohibit Ack_Enable
XWIN1
(fixed)
Offset
imaging
RAW Preview
RAW X-Ray image
(If HW-offset is disabled)
RAW offset(Or corrected)
image
1. Send Cmd“PrepAcq” on UI “Acquire” page.
2. After receiving the Cmd_PrepAcq, it will start the prepare process, and send back the
acknowlage of “Prohibit” and “Enable”, the “XWIN” will be started.
3. The XWIN is configured by parameter “Clear Acq Delay Time” on “SDK” page, the unit is “ms”.
4. User needs to make sure the X-Ray ends within the XWIN.
5. The detector will send the images after the XWIN closed.
6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.2.3. Work Flow(Prep+Acq)