User's Manual
1. Safety Information
User Manual of NDT1013LA 15
⚫ The product should be warmed up for 15 minutes before exposure or updating the gain map
and defect map.
⚫ Make sure wave form of the energy going to the X ray tube is square not pulse.
⚫ Be cautious with circumstance that someone has radio isotope recently injected into them, it
may cause panel transmit image without x ray.
⚫ Once powered off, please wait at least 60s before power on again
During exposure
⚫ Do not move Power Cable or Ethernet Cable during exposure, or it may cause image noise or
artifacts, even incorrect images.
⚫ Do not use the product near the equipment generating a strong magnetic field. Otherwise, it
may cause image noise, artifacts or even incorrect images.
After Usage
⚫ After every examination, wipe the patient contact surfaces with disinfectants such as ethanol,
to prevent the risk of infection. For details on how to sterilize, consult a specialist.
⚫ Do not spray the product directly with disinfectants or detergents.
⚫ Wipe it with a cloth slightly damped with a neutral detergent. Do not use solvents such as
alcohol, benzene and acid. Doing so may damage the surface of the product.
⚫ It’s recommended to use a waterproof non-woven cover as the isolated layer between
product and the blooding patient.