User's Manual

Signal Input/Output Parts Port
Emissions Test EMC Standard
Test Levels
Professional healthcare facility environment
Electrostatic Discharge IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transients/burst IEC 61000-4-4
±1 kV
100 kHz repetition frequency
Conducted disturbances
induced by RF elds
IEC 61000-4-6
3 V, 0.15 MHz–80 MHz
6 V in ISM bands between 0.15 MHz and 80
MHz
80 % AM at 1 kHz
Reference Cables Provided Against EMC
Cable
Recommended
Cable Length
Shielded or
Unshielded
Number Cable Classication
LAN Cable
(conguration mode)
3.5 m Shielded 1 pcs Signal
Important Information Regarding Electromagnetic Compatibility (EMC)
Focus HD 43 should be used in the hospital environment except for near active HF
surgical equipment and the RF shielded room of a medical equipment system for
magnetic resonance imaging, where the intensity of electromagnetic disturbances is
high.
Use of Focus HD 43 adjacent or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify correct operation.
Focus HD 43 requires special precautions regarding EMC and needs to be installed only
by Carestream or authorized personnel and put into service according to EMC
information provided in the user manual.
Safety and Regulatory Information
AL4609 1-6 2021-09-21