User's Manual

Table Of Contents
Safety and Regulatory Information
AJ4310 | 2019-08-13 1–19
Emissions and Immunity Compliance to the
IEC60601-1-2 Standard
Electromagnetic Emissions
Electromagnetic Immunity
Emissions Test Compliance Electromagnetic Environment
RF emissions CISPR 11
Group 1, Class B
The TRIMAX 35C Detector uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby elec-
tronic equipment.
Harmonic distortion IEC 61000-3-2
Class A
The TRIMAX 35C Detector is suitable for
use in all establishments, including domes-
tic establishments and those directly con-
nected to the public low voltage power
supply network that supplies buildings used
for domestic purposes.
Voltage fluctuations and flicker IEC 61000-3-3
Compliance
Emissions Test EMC Standard
Test Levels
Professional healthcare facility environ-
ment
Electrostatic discharge IEC 61000-4-2 ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM field IEC 61000-4-3 3 V/m
80 MHz–2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3 Refer to Proximity Fields From RF Wireless
Communications Equipment
Rated power frequency mag-
netic fields
IEC 61000-4-8 30 A/m
50 Hz or 60 Hz