User's Manual

Table Of Contents
Safety and Regulatory Information
AJ4309 | 2019-08-13 1–19
Emissions and Immunity Compliance to the
IEC60601-1-2 Standard
Electromagnetic Emissions
Electromagnetic Immunity
Emissions Test Compliance Electromagnetic Environment
RF emissions CISPR 11
Group 1, Class B
The Carestream Focus 35C Detector uses RF
energy only for its internal function. There-
fore, its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
Harmonic distortion IEC 61000-3-2
Class A
The Carestream Focus 35C Detector is suit-
able for use in all establishments, including
domestic establishments and those directly
connected to the public low voltage power
supply network that supplies buildings used
for domestic purposes.
Voltage fluctuations and flicker IEC 61000-3-3
Compliance
Emissions Test EMC Standard
Test Levels
Professional healthcare facility environ-
ment
Electrostatic discharge IEC 61000-4-2 ±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM field IEC 61000-4-3 3 V/m
80 MHz–2.7 GHz
80 % AM at 1 kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3 Refer to Proximity Fields From RF Wireless
Communications Equipment
Rated power frequency mag-
netic fields
IEC 61000-4-8 30 A/m
50 Hz or 60 Hz