Mars1417V Digital Flat Panel Detector User Manual 4.6.1. Defect Template Check Click “LoadFile” “LocalFile” page on Choose the specified defect template, and click “Open” The defect template showed on the UI 64 will ................................................................................................................................................iRay Technology Co. Ltd.
Operation 4.6.2. Defect Template Modification Open the template specified defect The defect management dialog box will be showed Find the pixel that needs to be managed, type the coordinate of the pixel and click “Add”, the information will be added to the template If click “Delete”, the information will be deleted Click “Save” iRay Technology Co. Ltd.……………………………………………………………………………………….………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual It is similar to manage the defect pixel, If user need to add the defect line, type the coordinate of the line and click “Add” If the information needs to be deleted, click “Delete” 4.7. Correction and Calibration Management 4.7.1. Correction and Calibration template synchronization Panel supports correction and calibration template storage.
Operation Click “Read Status” If the activity shows disenable, please select the one and click “Active” to enable it 4.7.2. Correction and Calibration management Panel supports two ways to do correction and calibration. Software Correction and Calibration defines the scenario that Workstation completes all correction and calibration. If panel complete all correction and calibration by itself, it is named as Hardware Correction and Calibration.
Mars1417V Digital Flat Panel Detector User Manual Choose “HWPostOffset”, “HWGain”, “HWDefect”, the hardware-based calibration is on Choose “SWPostOffset”, “SWGain”, “SWDefect”, the software-based calibration is on 4.8. Firmware Update Panel supports the use of the Web way to upgrade the firmware, if a user needs to update the firmware, please complete the following steps. Before update, please make sure that the battery capacity is more than 80% or the DC power is connected. 4.8.1.
Operation User name: admin Password: iray Click “login” Click “ ” Click “Upgrade” Click “Browse” iRay Technology Co. Ltd.……………………………………………………………………………………….………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual Find the MCU file, select it and open it The selected file name will be displayed on the interface. Click “Upgrade” Click “close” There will show the progress bar. If upgrade success, the interface as shown in the picture on the right will be displayed. Otherwise, it means failure. 70 ................................................................................................................................................iRay Technology Co. Ltd.
Operation 4.8.2. FPGA Update Connect the panel to PC / Open a browser, and type “192.168.8.8” in the search bar, then click the “Enter” button User name: admin Password: iray Click “login” Click “ ” Click “Upgrade” iRay Technology Co. Ltd.……………………………………………………………………………………….………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual Click “Browse” Find the FPGA file, select it and open it The selected file name will be displayed on the interface. Click “Upgrade” Click “close” There will show the progress bar. 72 ................................................................................................................................................iRay Technology Co. Ltd.
Operation If upgrade success, the interface as shown in the picture on the right will be displayed. Otherwise, it means failure. 4.8.3. ARM Update Connect the panel to PC / Open a browser, and type “192.168.8.8” in the search bar, then click the “Enter” button User name: admin Password: iray Click “login” iRay Technology Co. Ltd.……………………………………………………………………………………….………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual Click “ ” Click “Upgrade” Click “Browse” Find the ARM file, select it and open it The selected file name will be displayed on the interface. Click “Upgrade” 74 ................................................................................................................................................iRay Technology Co. Ltd.
Operation Click “close” There will show the progress bar. If upgrade success, the interface as shown in the picture on the right will be displayed. Otherwise, it means failure. 4.8.4. ALL FIRMWARE Update All of the firmware (MCU, FPGA, and ARM) can be upgraded at the same time, if the file selected is like “Mars1417V3TIS_IMAGE_41_ALL_2019_04_15.ifm”. And the upgrade steps please refer the steps above. 4.9.
Mars1417V Digital Flat Panel Detector User Manual 4.10. Software 4.10.1. Main GUI iRay provides test tools, such as iDetector for testing the basic performance of detector. It can connect the detector, acquire image, image correct and calibrate.
Operation 4.10.2. Message Box 4.10.2.1. Status Box Status box defines the current status of panel. SN Serial Number of the detector Status Status of the detector, busy or ready Task The current task being executed Message Information Remaining power of the battery, showed as percentage 4.10.2.2. Progress Bar Progress Bar defines as following. If progress bar is Green when shooting X ray, image quality is acceptable, otherwise image quality would degrade. iRay Technology Co. Ltd.
Mars1417V Digital Flat Panel Detector User Manual 4.10.3. Configuration GUI 4.10.3.1. General Settings Except the following parameters, the value should not be modified for other parameters.
Operation Prep CapMode Reserved Yes Self CapEnable Reserved YES Self Cap Span Time Should not be modified, and keep the original value YES Trigger Mode Trigger mode YES Sequence Interval Time Should not be modified, and keep the original value YES Set Delay Time Exposure window for Freesync mode YES Exp Window Time Exposure Window for Software/Inner mode, the value YES should not be large than 10s Acquire Delay Time Reserved YES Integrate Time Should not be modified, and keep the or
Mars1417V Digital Flat Panel Detector User Manual 4.10.3.2.
Operation 4.10.3.3.
Mars1417V Digital Flat Panel Detector User Manual 5. REGULATORY INFORMATION................................................................................... 83 5.1. Medical equipment safety standards........................................................................................ 83 5.2. The compliance for each EMISSIONS and IMMUNITY standard or test specified by IEC60601-1-2 standard.................................................................................................................
Regulatory Information 5. Regulatory Information 5.1.
Mars1417V Digital Flat Panel Detector User Manual 009/(R)2012+A2:2010/(R)2012 MOD) CAN/CSA-C22.2 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance No.60601-1:14 ISO 15223-1:2016/ 15223-1:2016 EN ISO Medical devices—Symbols to be used with medical device labels, labeling and information to be supplied—Part 1: General requirements 5.2.
Regulatory Information 450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m 704-787 Pulse modulation 217Hz, 9V/m 800-960 Pulse modulation 18Hz, 28V/m 1700-1990 Pulse modulation 217Hz, 28V/m 2400-2570 Pulse modulation 217Hz, 28V/m 5100-5800 Pulse modulation 217Hz, 9V/m 710 745 780 810 870 930 1720 1845 1970 2450 5240 5500 5785 Input AC power Port Basic EMC standard Phenomenon Electrical transients/burst fast Surges Line-to-line Surges Line-to-ground Conducted disturbances by RF fields IEC 610
Mars1417V Digital Flat Panel Detector User Manual Input DC power Port Basic EMC standard Phenomenon Immunity test levels Professional healthcare facility environment Electrical transients/burst Conducted disturbances by RF fields fast IEC 61000-4-4 ±2 kV 100kHz repetition frequency 3V, 0.15MHz-80MHz induced IEC 61000-4-6 6V in ISM bands between 0.
Regulatory Information communications such as mobile (cellular) telephones. Electromagnetic interference may result in incorrect operation of the system and create a potentially unsafe situation. The minimum distance between the panel and other equipment should be larger than 12 inch. Mars1417V conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
Mars1417V Digital Flat Panel Detector User Manual Increase the separation between the panel and receiver. Connect the panel into an outlet different from the receiver is connected. Consult the distributor or an experienced radio/TV technician for help. 5.4. Battery Safety Standards Standards Description CAN/CSA E62133:13 1st Ed. Rev.
Trouble Shooting 6. TROUBLE SHOOTING.................................................................................................90 iRay Technology Co. Ltd.………………………………………………………………………………………………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual 6. Trouble Shooting Please refer to service manual. If the problem persists, turn off the panel and contact iRay service department (service@iraygroup.com). We would provide the best service. 90 ................................................................................................................................................iRay Technology Co. Ltd.
Information 7. Service SERVICE INFORMATION.......................................................................................... 92 7.1. Product Lifetime............................................................................................................................. 92 7.2. Regular Inspection and Maintenance....................................................................................... 92 7.3. Repair.................................................................................
Mars1417V Digital Flat Panel Detector User Manual 7. Service Information 7.1. Product Lifetime The estimated product lifetime is up to 7 years under appropriate regular inspection and maintenance (battery 5 years). 7.2. Regular Inspection and Maintenance In order to ensure the safety of patients and operator, to maintain the performance and reliability of the panel, be sure to perform regular inspection at least once a year.
Appendix Appendix APPENDIX A INFORMATION OF MANUFACTURES...........................................................94 APPENDIX B INFORMATION OF MEDICAL DEVICE DIRECTIVE EUROPEAN REPRESENTATIVE..................................................................................................... 95 iRay Technology Co. Ltd.………………………………………………………………………………………………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual Appendix A Information of Manufactures Company: iRay Technology Taicang, Ltd. ADDRESS: NO.33 Xinggang Road, Taicang Port Economic and Technological Development Zone, Jiangsu, China ZIPCODE: 215434 TELEPHONE: +86-0512-53690872 94 ................................................................................................................................................iRay Technology Co. Ltd.
Appendix Appendix B Information of Medical Device Directive European Representative iRay Europe GmbH Address: In den Dorfwiesen 14, 71720 Oberstenfeld Germany Tel: +49-7062-977 88 00 Fax: +49-7062-976 0571 iRay Technology Co. Ltd.………………………………………………………………………………………………………………… …………….………..
Mars1417V Digital Flat Panel Detector User Manual FCC Regulations: Contains module’s FCC ID:2ACHK-01070189 This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Service Information This is because the device is designed to operate at multiple power levels so as to use only the power required to reach the network. In general, the closer you are to a wireless Base station antenna, the lower the power output. The exposure standard for wireless devices employing a unit of measurement is known as the Specific Absorption Rate, or SAR. The SAR limit recommended by the ICNIRP used by the general public is 2.