User's Manual

Operation
iRay Technology Co.
Ltd.……………………………………………………………………………………….…………………………
…………….………..
57
Calibrate module related
setting, such as making
correction and calibration
template, template in panel
could be uploaded to
workstation, and template in
workstation could also be
downloaded to panel.
Local File module related
setting, such as import Raw
or DCM image.
4.4. Correction and Calibration Template Generation
The correction and calibration should be performed after installation and it is recommended to perform
the new correction and calibration after any major change on the system settings and hardware
configuration. On the other hand, it is also recommended to do the correction and calibration in each 6
months.
4.4.1. Pre-offset Template Generation
If panel is configured to do Pre-offset correction, Pre-offset Template is necessary. See below