Please ch arge battery with charger following GB
9706.1 Standards provided by us
Do not mix battery with ones not provided by our
company
Do not charge battery with broken charger.
Substitution of battery inside main unit must be
carried out by qualified people
Do not touch output connector for adaptor
Do not remove the battery when the panel is powered
on only with the battery.
1.2. Notes for Using
When using the product, take the following precautions. Otherwise, problems may
occur and the product may not function correctly.
Before exposure
Be sure to check the product daily and confirm it work properly.
Sudden heating of the room in cold areas will cause condensation on the
product. In this case, wait until the condensation evaporates before performing
an exposure. If it is used when condensation is formed, problems may occur in
the quality of captured images. When an air-conditioner is used, be sure to
raise/lower the temperature gradually to prevent condensation.
The product should be warmed up for 15 minutes before exposure or updating
the gain map and defect map.
Make sure exposure rate is over 900nGy/s @70KV.
Make sure wave form of the energy going to the X ray tube is square not pulse.
Be cautious with circumstance that someone has radio isotope recently injected
into them, it may cause that panel transmit image without x ray.
During exposure
Do not move Power Cable, or it may cause image noise or artifacts, even
incorrect images.
Do not use the product near the equipment generating a strong magnetic field.
Otherwise, it may cause image noise, artifacts or even incorrect images.
After usage