User's Manual

Regulatory Information
KONICA MINOLTA,
INC..……………..…………………………………………………………………………………………………
……………….………..
83
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Signal input/output parts Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility
environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Electrical fast
transients/burst
IEC 61000-4-4
±1 kV
100kHz repetition frequency
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
The following shows information on reference cables provided against EMC
Cable
Recommended
cable length
Shielded or
Unshielded
Number Cable classification
AC Power Cable 3m Unshielded 1 pcs AC Power
DC Power Cable 3.5m Unshielded 1 pcs DC Power
LAN Cable
(configuration mode)
3m Shielded 1 pcs Signal
Important information regarding Electromagnetic Compatibility (EMC)
SKR 4000 requires special precautions regarding EMC and needs to be installed only by KONICA or
authorized personnel and put into service according to EMC information provided in the user manual.
SKR 4000 in use may be susceptible to electromagnetic interference from portable and mobile RF
communications such as mobile (cellular) telephones. Electromagnetic interference may result in
incorrect operation of the system and create a potentially unsafe situation.
SKR 4000 conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed: