User's Manual
Operation
KONICA MINOLTA,
INC..………….……………………………………………………………………………………………………
….…………….………..
51
4.1.2.2. Work Flow
Click “Prepare”
Panel preparation
Shoot X ray
X ray sensor in panel would trigger panel to start image acquisition.
Wait for image uploaded
4.1.2.3. Timing Setting
To set a clear scenario for program, see diagram below for details
Workstation
Panel
X Ray
Generator
1
2
X
ray
34
5
1. Workstation receives “prep” request.
2. X ray generator is ready for X ray shooting and begins to release X ray.
3. Panel starts uploading Pre-dark image and Light image to Workstation for preview. If hardware
offset is selected, panel would do offset first, and then upload preview image.
4. Panel starts uploading Post-dark image to Workstation. If hardware offset is chosen, panel would
do correction and calibration first, then upload processed image to Workstation.
5. Workstation into exposure prohibit state.