6. Regulatory Information
92 iRay Technology Co. Ltd..
standard
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5
±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz
80%AM at 1kHz
Voltage dips
IEC 61000-4-11
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at
Voltage
interruptions
IEC 61000-4-11
0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF communication
device and detector:
Portable RF communications equipment, including antennas, can effect medical electrical
equipment. The warning should include a use distance such as “be used no closer than 30 cm
(12 inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables specified
by the manufacturer
The following shows information on reference cables provided against EMC
Cable
Recommende
d
cable length
Shielded or
Unshielded
Number
Cable classification
AC Power Cable
3m
Unshielded
1 pcs
AC Power