iRay Technology Co. Ltd. 91
80% AM at 1kHz
Near fields from RF
wireless
communications
equipment
IEC 61000-4-3
Refer to table “Near fields from RF wireless
communications equipment”
Rated power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m
5240
5100-5800
Pulse modulation 217Hz, 9V/m
5500
5785
Input a.c. power port
Phenomenon
Basic EMC
Immunity test levels