Mars1717V Digital Flat Panel Detector User Manual
80 iRay Technology Co. Ltd..
5.2.2 Gain Calibration Template Generation
If the relative position between tube and detector changed or KV value changed, it
suggest to create gain template file.
Enter Create Gain page
Click "Start" button to start
process.
Click PREP button, acquire
image. Please exposure
after Acquire button enable.
And click Acquire button to
acquire image after
exposure end. Click Accept
button after acquired image.
If Current Value textbox is
yellow, click PREP button.
Re-acquire images after
adjust generator parameters.
Note: In different trigger
mode, the operation maybe
have little difference. Please
follow the UI tips.