5. Operation
iRay Technology Co. Ltd. 73
3. Panel finishes “Clear” and send a message to workstation “Exposure Enable”.
4. Workstation shows “Exposure Enable” on the IDetector’s message bar to tell
user shoot X-ray now.
5. User triggers X-ray generator to initialize and do anode rotation to prepare for
X-ray shooting.
6. X-ray generator finishes preparation for X-ray shooting and reminds user to
shoot.
7. X-ray generator starts releasing X-ray.
8. X-ray generator finishes x-ray shooting.
9. Workstation receives “Acquire” request, send command “Data Acquisition”
to panel.
10. Panel receives “Data Acquisition” from workstation, start data acquisition
operation.
11. Panel completes image acquisition and begins to send data to workstation.
12. Workstation receives all image data from panel
If Hardware Pre-offset and Hardware calibration is selected, image is the final
image.
If Hardware Post offset and Hardware calibration is selected, image got would be
preview image. After step12, panel would do another dark image acquisition.
With both light image and dark image, panel completes all the correction and
calibration process. Finally, panel uploads processed image to workstation.