User's Manual

Wireless Digital Flat Panel Detector Mars1717X
80 User Manual of Mars1717X
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Input a.c. power port
Phenomenon Basic EMC standard
Immunity test levels
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
Surges
Line-to-ground
IEC 61000-4-5
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz
80%AM at 1kHz
Voltage dips IEC 61000-4-11
0% UT; 0.5 cycle
At 0 , 45 , 90 , 135 , 180 , 225 , 270 and 315
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0
Voltage interruptions IEC 61000-4-11 0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF communication
device and detector:
Portable RF communications equipment, including antennas, can effect medical electrical