4.10.1. HW pre-offset Template Generation
Enter Acquire interface, select
HWPostOffset option
Enter Calibrate interface, click
UpdateHWPreOffset button.
Waiting until status bar displayed:
"Task succeed:
4.10.2. Gain Calibration Template Generation
If the relative position between tube and detector changed or KV value changed, it suggest to
create gain template file.
Enter Create Gain page
Click "Start" button to start
process.