User's Manual

Wireless Digital Flat Panel Detector Mars1717X
64 User Manual of Mars1717X
Operation Instructions for Image Acquisition
Mars1717X provides SDK for users to integrate detector into their DR system. Additionally, it also
provides an application for demonstration, i.e. IDetector. User can use IDetector to control detector
without DR system.
4.6. Steps for acquiring image
Make sure the hardware is connected correctly and then power on.
Once powered off, please wait at least 60s before power on again
Wait until initialization is complete
Connect the software
choose the synchronization mode
Generate HWPreOffset, Gain and Defect template after the detector reaches thermal
equilibrium
Acquire images in the selected mode
To Acquire X-ray image is the main operation of Mars1717X. Most importantly, detector should
build synchronization with X-ray generator. Mars1717X has one synchronization modes to acquire
X-ray image, which is Software Mode.
4.7. Software Mode
4.7.1. Block Diagram
Software mode is the basic way to acquire X-ray image. Please see figure below for general feature.
Workstation is a host PC device installed with iDetector and SDK. FPD is the Flat Panel Detector
and HVG is the High Voltage Generator. In this mode, Workstation does not have to control X-ray
generator. Users would decide when to shoot X-ray.
HVG
X-Tube
FPD
Work
Station