User's Manual

1. Safety Information
User Manual of Mars1717X 17
The product should be warmed up for 15 minutes before exposure or updating the gain map
and defect map.
Make sure wave form of the energy going to the X ray tube is square not pulse.
Be cautious with circumstance that someone has radio isotope recently injected into them, it
may cause panel transmit image without x ray.
Once powered off, please wait at least 60s before power on again
During exposure
Do not move Power Cable or Ethernet Cable during exposure, or it may cause image noise or
artifacts, even incorrect images.
Do not use the product near the equipment generating a strong magnetic field. Otherwise, it
may cause image noise, artifacts or even incorrect images.
After Usage
After every examination, wipe the patient contact surfaces with disinfectants such as ethanol,
to prevent the risk of infection. For details on how to sterilize, consult a specialist.
Do not spray the product directly with disinfectants or detergents.
Wipe it with a cloth slightly damped with a neutral detergent. Do not use solvents such as
alcohol, benzene and acid. Doing so may damage the surface of the product.
-woven cover as the isolated layer between
product and the blooding patient.