5.2.3. Work Flow(Prep+Acq)
IDLE Prepare
XWIN2
(dynamic)
X-Ray
imaging
Cmd_Prep Ack_Prohibit Ack_Enable
XWIN2
(dynamic)
Offset
imaging
RAW Preview
RAW X-Ray image
(If HW-offset is disabled)
RAW offset(Or corrected)
image
Cmd_SingleAcq
X-Ray
1. Send Cmd“Prep” on UI “Acquire” page.
2. After receiving the Cmd_Prep, it will start the prepare process, and send back the acknowlage
of “Prohibit” and “Enable”, the “XWIN” will be started.
3. The max XWIN is configured by parameter “Exp Window Time” on “Detector” page
“Parameter” tab, the unit is “ms”.
4. User starts the X-Ray.
5. Send “SingleAcq” on UI “Acquire” page after the X-Ray is end.
6. The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.3. AED Mode
5.3.1. Inner
IDLE
XWIN1
(fixed)
X-Ray
imaging
XWIN1
(fixed)
Offset
imaging
X-Ray Info
RAW Preview
RAW X-Ray image
If HW-offset is disabled
RAW Offset(Or corrected)
image
Low power
Prep Ack
1. The detector is in low power state, user needs to send Cmd “Prep” to make the detector exit
to idle state which indicated by the acknowledge to Cmd “Prep”.
2. When the detector is in idle state, user can start the X-Ray any time.