Update the Firmware
The Internet can be used to upgrade the detector rmware.
This procedure applies to MCU, FPGA, and ARM.
1. Connect the panel to a PC.
2. Open a browser, type detector IP in the search bar, and then press Enter.
3. Enter the following and select login.
• User name: admin
• Password: admin
4. Click on the three squares located on the left side of the screen.
5. Select Upgrade and then Browse.
6. Select one the rmware les from the list and select Open.
• 1717V3TISA07_IMAGE_MCU_xxxx_xx_13.ifm
• 1717V3TISA07_FPGA_xxxx_xx_13.ifm
• 1717V3TISA07_IMAGE_ARM_xxxx_xx_13.ifm
Important:
Selecting 1717V3TISA07_IMAGE_ALL_xxxx_xx_13.ifm will allow all three versions of
rmware to be updated at the same time.
7. When the selected le name is displayed on the interface, select Upgrade.
8. Select Close when the Notice box is displayed.
The progress bar will be displayed.
9. If the upgrade is successful, the following interface screen will be displayed.
Note:
If the screen is not displayed, the upgrade has failed.