Correction and Calibration Management
Correction and Calibration Template Synchronization
The detector supports correction and calibration template storage. Templates in the
detector can be uploaded to a workstation, and templates in a workstation can be
downloaded to a detector.
1. After generating the offset, gain, and defect templates, select the templates and
select Download to FPD.
2. Select Download.
3. Select Read Status.
4. If the Activity column shows disable, select the row and select Active to enable it.
Correction and Calibration Management
The detector supports the following ways to do correction and calibration:
• In software correction and calibration, the workstation completes all correction and
calibration.
• In hardware correction and calibration, the panel completes all correction and
calibration.
1. User can set the calibration method on the Detector page.
Note:
The hardware-based calibration is on.
2. Select:
• HWPostOffset
• HWGain
• HWDefect
Note:
The hardware-based calibration is on.
3. Select:
• SWPostOffset
• SWGain
• SWDefect
Note:
The software-based calibration is on.