User's Manual

1. On the Defect Calibration page, start the exposure. Eight images need to be
captured.
2. Select Start.
3. Select PREP and start the exposure.
4. When completed, select Acquire to get the light image.
5. If the value meets the expected value, select Accept, and then acquire the other
seven images.
If the value does not meet the expected value, do not select Acquire. Adjust the
exposure dose, then select PREP to acquire the light image again.
6. When all eight images are created, select Generate to generate the gain template.
Note:
Make sure your x-ray dose is correct. If your dose is out of the range, iDetector will remind
you to adjust the dose. Then you can select start creating and try again.
If users operate with two panels, SDK has a probability of automatically quitting.