Correction and Calibration Template Generation
Correction and calibration should be performed after installation and every six months.
The new correction and calibration should be performed after any major change on
the system settings and hardware conguration.
Pre-offset Template Generation
If panel is congured to do Pre-offset correction, Pre-offset Template is necessary.
1. Select Calibrate.
2. Select Start Generate Templates.
3. Select Create Offset.
4. Select Start create offset template le.
The screen will display Offset Map Generating.
5. When complete, the screen will display Offset MAP Generated!
Gain Calibration Template Generation
Before doing this procedure, make sure SID1.2m, no copper is required.
Actual screens may be different from those shown in this procedure.
Note:
Use software post offset correction.
1. On the gain template generating page, ve images need to be created.
2. Select Start.
3. Select PREP and start the exposure.
4. When completed, select Acquire to get the light image.
5. If the value meets the expected value, select Accept, and then acquire the other
four images.
If the value does not meet the expected value, do not click Accept. Adjust the
exposure dose, and then click PREP to acquire the light image again.
6. When all ve images are created, select Generate to generate the gain template.
Defect Correction Template Generation
Before doing this procedure, make sure SID1.2m, no copper is required.
Actual screens may be different from those shown in this procedure.
Note:
Use software post offset mode.