Establish a Connection with the Detector
1. Open SDK, select the detector, and select Connect.
2. Conrm that the values for IP address and Port are the same as the values in
cong.ini.
Note:
The value for Cfg_HostPort should be the default of 28000.
Note:
When the connection is changed to a different network card, the user must reconnect the
detector using a different IP address.
The rule of Multi-Share control is based on the IP address. The second terminal with a
different IP address is not allowed to operate a detector after the rst one is connected. If
there is no command transmission between the detector and workstation after 5 minutes,
the detector releases access authority.
Appendix (For Service Personnel Only)
AJ4312 | 2020-02-10 XIII