15. Select:
• Detector
• Wi
• Read Cong
• Client
16. Select Add and enter the SSID and Password. Select Apply.
17. Choose SSID and select the one with a check mark. Select Write Cong to save the
parameters.
18. Turn on the wireless router.
a. Make sure there is a wired connection between the outer, work station, and IP
10.0.1.251.
b. Select Read Wi Status to check wireless transmission status, numerical value
occurred means the link is up and available.
The detector will connect to the wireless AP the next time it is powered on.
Wireless AP Mode
A wired cable can also be used to congure detector in wireless client mode. The wired
connection should only be used by the service operator.To start conguration with
wired cable, it is necessary to complete this procedure.
1. Connect the panel to the workstation with the Ethernet cable.
2. Select the Detector tab and then Wi.
3. In the Mode eld, select AP.
4. Select Read Cong to display the default settings.
5. Change the settings for SSID and password.
Important:
Make sure that the SSID is different from others already used.
6. In the Frequency eld, click on the arrow and select a value from the drop-down
list..
7. In the Country eld, click on the arrow and select a country from the drop-down
list.
8. In the Channel eld, click on the arrow and select a clean frequency and channel.
9. Select Write Cong to save the settings.
Note:
Do not remove the wired cable until the FPD status is Ready.
Appendix (For Service Personnel Only)