mode, the user does not interact with the workstation. After taking x-rays, images
immediately appear on the screen.
Work Flow
1. X-ray generator is ready to take x-rays and starts releasing the x-ray.
2. Workstation receives Exposure Prohibited from the panel.
3. Panel starts uploading a Pre-dark image and a Light image to the workstation for
preview. If hardware offset is selected, panel rst performs an offset and then
uploads the preview image.
4. Panel starts uploading Post-dark image to the workstation. If hardware offset is
selected, panel rst performs correction and calibration and then uploads the
processed image to the workstation.
5. Workstation receives Exposure Enable from the panel.