User's Manual
Table Of Contents
- Contents
- Notices and Conventions
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- Appendix (For Service Personnel Only)
- Software and Settings
- Operating Modes
- Software Installation
- Set the Connection Mode
- Shortcuts
- Establish a Connection with the Detector
- Configure the Detector
- Correction and Calibration Template Generation
- Image Check and Upload
- Defect Template Check and Modification
- Correction and Calibration Management
- Update the Firmware
- Publication History
Operating Modes
Software Mode
Block Diagram
Software mode is the basic way to acquire an x-ray image.
The workstation is a host device installed with iDetector and SDK. Installation describes
how to establish connections between the panel and workstation. In software mode,
the workstation does not control the x-ray generator. Users decide when to take x-rays.
Work Flow
1. Workstation receives prep request and sends clear command to the panel.
2. Panel receives clear from the workstation and starts clearing leakage from the
panel. Meanwhile, the panel sends an Exposure Prohibited message to the
workstation.
3. Panel nishes clear and sends an Exposure Enable message to the workstation.
4. Workstation shows Exposure Enable on the iDetector’s message bar to tell the
user to take the x-ray now.
5. User triggers the x-ray generator to initialize and do an anode rotation to prepare
for taking x-rays.
6. X-ray generator nishes preparation for taking x-rays and reminds the user to take
the x-ray.
7. X-ray generator starts releasing the x-ray.
8. X-ray generator nishes taking the x-ray.
9. Workstation receives acquire request and sends Data Acquisition command to
the panel.
10. Panel receives Data Acquisition from the workstation and starts data acquisition
operation.
11. Panel completes image acquisition and begins to send data to the workstation.
12. Workstation receives all image data from the panel after calibration if hardware
calibration is on.
Time Setting
To set a clear scenario for programming, see the diagram below.
Appendix (For Service Personnel Only)