- Publication History
Temperature Temperature
Change
Humidity Atmospheric
Pressure
Pressure Change
TRIMAX 43 detectors shall operate at a specied altitude of not more than 3000.0 m (9842.5 ft). The
environment specic is only for the detector.
IT Network
Purpose for IT-network
Transmission of image data and command/status communication between the detector
and the workstation.
Required Features
Wireless communication follows IEEE 802.11a/b/g/n/ac protocol. It works on 2.4 GHz
and 5 GHz.
It supports at least 2 routers.
Hazardous Situations from Failure of the IT Network
• Incompatibility of the operating system is not compatible
• Inability to update or change software
• Incompatibility of the interface
• Inconsistency of interface or format leads to data distortion
• Data transfer protocol error
• Data output failure
Required Conguration
The wireless card and the detector must work on the same IP segment such as
192.168.8.XXX.
They must support IEEE 802.11a/b/g/n/ac.
Technical Specications
Item Specications
Wireless Standard IEEE 802.11a/b/g/n/ac
Frequency Range
2.412-2.472 GHz: ch1-ch13
5.18-5.24 GHz: ch36-ch48
5.745-5.85 GHz: ch149-ch165
Overview
AJ4312 | 2020-02-10 2-11