Emissions Test EMC Standard
Test Levels
Professional healthcare facility environment
Surges
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1kV
Surges
Line-to-ground
IEC 61000-4-5 ±0.5 kV, ±1kV, ±2 kV
Conducted disturbances
induced by RF elds
IEC 61000-4-6
3 V, 0.15 MHz–80MHz
6 V in ISM bands between 0.15 MHz and
80 MHz
80 % AM at 1 kHz
Voltage dips
IEC 61000-4-11
0 % UT; 0.5 cycle
At 0 º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º and
315 º
Voltage dips IEC 61000-4-11
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0 º
Voltage interruptions IEC 61000-4-11 0 % UT; 250/300 cycles
Signal Input/Output Parts Port
Emissions Test EMC Standard
Test Levels
Professional healthcare facility environment
Electrostatic Discharge IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Electrical fast transients/burst IEC 61000-4-4
±1 kV
100 kHz repetition frequency
Safety and Regulatory Information
AJ4312 | 2020-02-10 1-19