User's Manual
Table Of Contents
- Contents
- Notices and Conventions
- 1 Safety and Regulatory Information
- 2 Overview
- 3 Installation
- 4 Operation
- Appendix (For Service Personnel Only)
- Software and Settings
- Operating Modes
- Software Installation
- Set the Connection Mode
- Shortcuts
- Establish a Connection with the Detector
- Configure the Detector
- Correction and Calibration Template Generation
- Image Check and Upload
- Defect Template Check and Modification
- Correction and Calibration Management
- Update the Firmware
- Publication History
Emissions and Immunity Compliance to the
IEC60601-1-2 Standard
Electromagnetic Emissions
Emissions Test Compliance Electromagnetic Environment
RF emissions
CISPR 11
Group 1, Class B
The TRIMAX 43C Detector uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
Harmonic distortion
IEC 61000-3-2
Class A
The TRIMAX 43C Detector is suitable for
use in all establishments, including
domestic establishments and those directly
connected to the public low voltage power
supply network that supplies buildings
used for domestic purposes.
Voltage uctuations and icker
IEC 61000-3-3
Compliance
Electromagnetic Immunity
Emissions Test EMC Standard
Test Levels
Professional healthcare facility
environment
Electrostatic discharge IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM eld IEC 61000-4-3
3 V/m
80 MHz–2.7 GHz
80 % AM at 1 kHz
Proximity elds from RF wireless
communications equipment
IEC 61000-4-3
Refer to Proximity Fields From RF Wireless
Communications Equipment
Rated power frequency
magnetic elds
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Safety and Regulatory Information
AJ4312 | 2020-02-10 1-17