6. Regulatory Information
68 iRay Technology Co., Ltd..
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5 ±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and 80MHz
80%AM at 1kHz
Voltage dips IEC 61000-4-11
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
Voltage
interruptions
IEC 61000-4-11 0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF communication device
and detector:
Portable RF communications equipment, including antennas, can affect medical electrical
equipment. The warning should include a use distance such as “be used no closer than 30 cm
(12 inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables specified by
the manufacturer”
Important information regarding Electromagnetic Compatibility (EMC)
Mars1717VS requires special precautions regarding EMC and needs to be installed only by iRay or
authorized personnel and put into service according to EMC information provided in the user manual.
Mars1717VS in use may be susceptible to electromagnetic interference from portable and mobile RF
communications such as mobile (cellular) telephones. Electromagnetic interference may result in
incorrect operation of the system and create a potentially unsafe situation.