User's Manual

Mars1717V Digital Flat Panel Detector User Manual
iRay Technology Co., Ltd. 67
Radiated RF EM
field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Near fields from RF
wireless
communications
equipment
IEC 61000-4-3
Refer to table “Near fields from RF wireless
communications equipment”
Rated power
frequency magnetic
fields
IEC 61000-4-8 30A/m
50Hz or 60Hz
Near fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine, 28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
Input a.c. power port