3) After exposure, click “Acquire” to get the light image;
4) If the value meets the expected value, click “Accept”, then get the other 11 images;
5) If the value does not meet the expected value, please do not click the “Accept”, and adjust the exposure
dose, then click the “PREP” to get light image again;
6) After getting 12 images, click “Generate” to generate gain template.
Notes: 1 please use hardware post offset correction.
5.2.3 Defect Correction Template Generation
The operation process is as follows:
1) On the “Defect Calibration” page, start exposure, there are 8 images need to be captured;
2) Click “Start” button;
3) Click “PREP”, and start exposure;
4) After exposure, click “Acquire” to get the light image;
5) If the value meets the expected value, click “Accept”, then get the other 7 images;
6) If the value does not meet the expected value, please do not click the “Accept”, and adjust the exposure
dose, then click the “PREP” to get light image again;
7) After getting 8 images, click “Generate” to generate gain template.
Note:
1. Please use hardware post offset mode.
2. Make sure your x ray dose is right, if your dose is out of the range, iDetector will remind you to adjust
the dose. Then you can click “start creating” and try again.
5.3 Local Image Check
“OPEN” provides two features for image check and uploading. Local Image Check,
Panel Image Upload. Local Image Check defines function to check image saved in
Workstation. Panel Image Upload defines function to upload images stored in panel.