5.2 Correction and Calibration Template Generation
The correction and calibration should be performed after installation and it is
recommended to perform the new correction and calibration after any major change
on the system settings and hardware configuration. On the other hand, it is also
recommended to do the correction and calibration in each 6 months.
5.2.1 Pre-offset Template Generation
If panel is configured to do Pre-offset correction, Pre-offset Template is necessary. See below
Select “Calibrate”
Click “Start Generate
Templates”
Click “UpdateHWPreOffset”
5.2.2 Gain Calibration Template Generation
On Gain template generating page, there are 12 images that need to be got.
The operation process is as follows:
1) Click “start” button;
2) Click “PREP”, and start exposure;