User's Manual

5. Operation
iRay Technology Co., Ltd. 55
5.1.2.3 Timing Setting
To set a clear scenario for programming, see diagram below for details
1. Workstation receives “prepacquire” request, send command “clear” to panel.
2. Panel receives “clear” from workstation, starts clearing leakage of panel.
Meanwhile, panel send a message to workstation “Exposure Prohibited”.
3. Panel finishes “Clear” and send a message to workstation “Exposure Enable”.
4. Workstation shows “Exposure Enable” on the IDetector’s message bar to tell
user shoot X-ray now.
5. User triggers X-ray generator to initialize and do anode rotation to prepare for
X-ray shooting.
6. X-ray generator finishes preparation for X-ray shooting and reminds user to
shoot.
7. X-ray generator starts releasing X-ray.
8. X-ray generator finishes x-ray shooting.
9. Time out, panel acquires data.