User's Manual
Mars1717V Digital Flat Panel Detector User Manual
54 iRay Technology Co., Ltd..
5.1.2 Software Mode
5.1.2.1 Block Diagram
Software mode is the basic way to acquire x-ray image. Please see figure below for
general feature
HVG
X-Tube
FPD
Work
Station
Workstation is a host device installed with iDetector and SDK. Chapter 3 has
described how to establish connection between panels and workstation. In software
mode, workstation does not control x-ray generator. Users would decide when to
shoot x-ray.
5.1.2.2 Work flow
Click “PrepAcq”
Panel preparation, Wait until the warning message change from “Exposure
Prohibit” to “Exposure Enable”
Shoot X-ray at any time. However the longer time you wait. The worse
image would be. So please shoot once X-ray generator is ready.
Wait for image uploaded