User's Manual

Mars1717VS Digital Flat Panel Detector User Manual
10 iRay Technology Co., Ltd..
1.1 Safety Precautions
Follow these safeguards and properly use the equipment to prevent injury and damage to
any equipment/data.
WARNING
Installation and
environment of
use
Do not use or store the equipment near flammable
chemicals such as alcohol, thinner, benzene, etc.
If chemicals are spilled or evaporate, it may result in fire or
electric shock through contact with electric parts inside the
equipment. Also, some disinfectants are flammable. Be
sure to take care when using them.
Do not connect the equipment with anything other than
specified.
Doing so may result in fire or electric shock.
All the patients with active implantable medical devices
should be kept away from the equipment.
Power supply
Do not operate the equipment using any type of power
supply other than the one indicated on the rating label.
Otherwise, it may result in fire or electric shock.
Do not handle the equipment with wet hands.
You may experience electric shock that could result in
death or serious injury.
Do not place heavy object such as medical equipment
on cables and cords. Do not pull, bend, bundle, or step
on them to prevent their sheath from being damaged,
and do not alter them neither.
Doing so may damage the cords which could result in fire
or electric shock.
Do not supply power to more than one piece of
equipment using the same AC outlet.
Doing so may result in fire or electric shock.
Do not turn ON the system power when condensation
has formed on the equipment.
Doing so may result in fire or electric shock.
Do not connect a multiple portable socket-outlet or
extension cord to the system.
Doing so may result in fire or electric shock.
To avoid the risk of electric shock, this equipment
must only be connected to power supply with
protective earth.
Not doing so may result in fire or electric shock.
Securely plug the power cord into the AC outlet.
If contact failure occurs, or if metal objects come into
contact with the exposed metal prongs of the plug, fire or
electric shock may result.
Be sure to turn OFF the power to each piece of
equipment before connecting or disconnecting the
cords.
Otherwise, you may get an electric shock that could result
in death or serious injury.
Be sure to hold the plug or connector to disconnect
the cord.