Wireless Digital Flat Panel Detector Mars1717VS User Manual Document Version: A0 Document ID: 119-201-01 Release Date: 2019.09.16 Before operating, please read this user manual and pay attention to all safety precautions. Please ensure that this user's manual is properly maintained so that it can be accessed at any time ( reserve ). Please use it correctly on the basis of full understanding of the content.
About FCC This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device must not cause harmful interference; (2) This device must accept any interference received, including interference that may cause undesired operation. Attention must be paid to the fact that changes or modifications not expressly approved by the party responsible for compliance can void the user’s authority to operate the equipment.
To Customers To Customers Congratulations on your purchase of the Fixed Digital Flat Panel (hereinafter referred to as Mars1717VS) which is manufactured by iRay Technology Co.Ltd. (Hereinafter referred to as iRay). At iRay, we strive to not only make the world-class products that deliver the best value possible to our customers but also offer the highest quality of service and customer care. Please take time to read through this user guide in order to utilize the product effectively.
Mars1717V Digital Flat Panel Detector User Manual Caring for your environment This symbol indicates that this product is not to be disposed of with your residential or commercial waste. Recycling iRay Equipment Please do not dispose of this product with your residential or commercial waste. Improper handling of this type of waste could have a negative impact on health and on the environment.
Mars1717VS Digital Flat Panel Detector User Manual Symbols and Conventions The following symbols and conventions are used throughout the user guide. This symbol is used to identify conditions under which improper use of the product may cause death or serious personal injury. This notice is used to identify conditions under which improper use of the product may cause minor personal injury. This notice is used to identify conditions under which improper use of the product may cause property damage.
To Customers Labels and markings on the equipment The contents of the labels and markings on iRay Mar1717VS product are indicated below: Symbol Guide Caution: please refer to the instructions in the user manual. This symbol is used to indicate that the equipment has passed CE testing and it is followed by the CE number. This symbol is used to identify the manufacture series number which is after, below or adjacent to the symbol.
Mars1717VS Digital Flat Panel Detector User Manual Package symbol, fragile. Package symbol, keep away from sunlight. Package symbol, keep dry. Package symbol, keep the equipment up right. Package symbol, do not roll the transportation package. Type-B applied part Package symbol, this symbol is used to indicate stacking limit number. ×3 6 IPX3 for working surface only iRay Technology Co., Ltd..
Mars1717VS Digital Flat Panel Detector User Manual Contents TO CUSTOMERS.......................................................................................................................... 2 CONTENTS .................................................................................................................................. 7 1. SAFETY ................................................................................................................................. 9 1.1 Safety Precautions .....
Contents 4.8 List of the HAZARDOUS SITUATIONS resulting from a failure of the IT-NETWORK........ 50 5. OPERATION ........................................................................................................................... 51 5.1 Main Operation ...................................................................................................................... 52 5.1.1 Steps for acquiring image ......................................................................................... 53 5.1.
Mars1717VS Digital Flat Panel Detector User Manual 1. Safety 1.1 Safety Precautions ................................................................................................................ 10 1.2 Notes for Using of FPD ....................................................................................................... 14 1.3 Notes for Using of Battery .................................................................................................... 14 9 iRay Technology Co., Ltd.
Mars1717VS Digital Flat Panel Detector User Manual 1.1 Safety Precautions Follow these safeguards and properly use the equipment to prevent injury and damage to any equipment/data. WARNING Installation and environment of Do not use or store the equipment near flammable chemicals such as alcohol, thinner, benzene, etc. If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the equipment. Also, some disinfectants are flammable.
1. Safety If you pull the cord, the core wire may be damaged, resulting in fire or electric shock. WARNING Handling Never disassemble or modify the equipment. No modification of this equipment is allowed. Parts of the Mar1717VS that are not serviced or maintained while in use with the patient. Doing so may result in fire or electric shock. Also, since the equipment incorporates parts that may cause electric shock as well as other hazardous parts, touching them may cause death or serious injury.
Mars1717VS Digital Flat Panel Detector User Manual CAUTION Installation and environment of use Do not install the equipment in any of the locations listed below. Doing so may result in failure, malfunction, equipment falling, fire or injury.
1. Safety CAUTION Handling Handle the equipment carefully. Do not submerge the equipment in water. The internal image sensor may be damaged if something hits against it or it is dropped. Do not place excessive weight on the equipment. Otherwise, the internal image sensor may be damaged and image may be incorrect. Uniform load:150 kg over the whole area of the detector surface. Local load:100 kg on an area 4 cm diameter.
Mars1717VS Digital Flat Panel Detector User Manual 1.2 Notes for Using of FPD When using the equipment, take the following precautions. Otherwise, problems may occur and the equipment may not function correctly. Before exposure Be sure to check the equipment daily and confirm that it works properly. Be sure there be a battery installing on the Mars1717VS to avoid the power off suddenly Sudden heating of the room in cold areas will cause condensation to form on the equipment.
1. Safety Battery is shipped in package without detector, remaining capacity should be charged greater than or equal to 20% but less than 30%, If storage without use, charge greater than or equal to 20% but less than 30% every 2 month, or it causes damage to battery. If battery remaining capacity is lower than 20%, delivery is not allowed, or it is possible to be over discharge. Battery storage and working environment should strictly follow specification.
Mars1717VS Digital Flat Panel Detector User Manual 2. General Description 2.1 Scope .................................................................................................................................... 17 2.2 Model ........................................................................................................... 错误!未定义书签。 2.3 Characteristic............................................................................................... 错误!未定义书签。 2.4 Intended Use ......................
2. General Description Mars1717VS is a cassette-size wireless X-ray flat panel detector based on amorphous silicon thin-film transistor technologies. It is developed to provide the good quality of radiographic image, which contains an active matrix of 3070×3070 with 139um pixel pitch. Panels’ scintillator is CsI (Cesium Iodide). Mars1717VS supports wireless communication between panel and Workstation, and can be used as a real portable panel. 2.
Mars1717V Digital Flat Panel Detector User Manual According to the Mars1717VS series INTENDED USE and the result of risk management, image acquisition and data transmission are defined as ESSENTIAL PERFORMANCE. Getting dark image proves that ESSENTIAL PERFORMANCE does not influence INTENDED USE. Method for getting dark image in detail refers to section “installation” and “operation” 2.
2. General Description 2.6 Product Components Mars1717VS comes with battery package. Once powered on, it would build a connection with Workstation through Ethernet cable (only for service) or Wireless connection. Item Detector Battery iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual 20 Ethernet Cable 1pcs (Only for service) 1m Ethernet Adapter 1pcs Battery Charger 1pcs iRay Technology Co., Ltd..
2. General Description 2.7 Components Description And Specification 2.7.1 Detector External Signals Input A B External Power Input No. Item iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual A Positive Pole NA B Negative Pole NA Indicator No. Item Description D Power E Link F Status Power button of control panel Link button of control panel Status button of control panel Detector Specification Item Specification Model Mars1717VS-VSI (CsI) Image Sensor a-Si (Amorphous Silicon) TFT Pixel Size 139μm Effective Array 3070 x 3070 Effective Area (H x V) 426.7mm x 426.7mm Gray scales 16bit Limiting Spatial Resolution 3.
2. General Description Weight (with one battery) ≤4.2 kg Image Transfer Wireless: IEEE802.11 a/b/g/n/ac Wireless Frequency Range 2.412~2.472GHz, 5.18~5.22GHz; 5.745~5.85GHz Data Transmission Power 13dBm (Typ.) @802.11a 16dBm (Typ.) @802.11b 14dBm (Typ.) @802.11g 13dBm (Typ.) @802.11n HT20 11dBm (Typ.) @802.11n HT40 16dBm@2.4GHz 13dBm@5.8GHz Wireless Modulation Wireless Band 11b: DSSS (DBPSK, DQPSK and CCK) 11a/g/n: OFDM (BPSK, QPSK,16QAM, 64QAM) 2.4GHz≤40MHz 5.19GHz≤40MHz 5.
Mars1717V Digital Flat Panel Detector User Manual 2.7.2 Battery Item Name Description A Battery Label / B Battery Interface 8 Pin Battery connector C Pilot Pin / D Indicator Installation direction indicator 24 iRay Technology Co., Ltd..
2. General Description Dimension and Specification Item Model Rated Capacity Nominal Voltage Charge Voltage Discharged End Voltage Charging Method Operating Temperature Storage Temperature Relative Humidity Dimension (L × W × H) Weight iRay Technology Co., Ltd. Specifications Battery-KV Typ. 4180mAh @ Discharge 0.2C 10.8V 12.6±0.05V 9V CC-CV Charge 0℃-+60℃, Discharge-10℃-+60℃ ≤3 month -20℃-+45℃ ≤6 month -20℃-+35℃ 5%~95% 210 x 115 x 7.5 mm 0.
Mars1717V Digital Flat Panel Detector User Manual 2.7.3 Battery Charger Item Name Description A Battery Interface 8 Pin Battery connector B Capacity Indicator The indicator definition is as follow C Power Indicator The indicator definition is as follow D Hand Pull Position / E The limit ball plug / F DC Jack 24V DC input Dimension and Specification 26 iRay Technology Co., Ltd..
2. General Description Item Specifications Model Charger-Combo Simultaneous Charging 1 battery packs Full charging time ≤3 hours Rated power supply 90~264V(AC) Dimension (L × W × H) 240 x 184 x 38 mm Weight 0.
Mars1717V Digital Flat Panel Detector User Manual 2.7.4 Power supply Mars1717VS supports both DC Power and Battery package input. Item Specifications DC Power 24V(DC), 0.75A Battery Package 10.8V(DC),1.5A 2.7.5 Recommended Application Condition Item Description Windows 7 Operating System 32/64bit CPU Memory Hard Disk Intel Core i7 3.6G 4G DDR3 640 G Intel Pro EXP9301CT PRO LAN Card Gigabit Network Adapter with PCIe interface 2.7.
Mars1717V Digital Flat Panel Detector User Manual 3. Installation 3.1Panel Installation .................................................................................................................... 30 3.1.1 Attach Battery Pack .................................................................................................... 30 3.1.2 Attach DC Power ........................................................................................................ 31 3.1.4 Indicator ...........................
Mars1717V Digital Flat Panel Detector User Manual 3.1 Panel Installation 3.1.1 Attach Battery Pack Mars1717VS can be powered by both battery package and DC power. Once battery package is inserted or DC power is on, Panel would be activated immediately. If none of battery and DC power is on, Mars1717VS would power off. Please see below for battery installation. Make sure that the connectors on the battery package are pointed to the cave in battery compartment.
3. Installation 3.1.2 Attach DC Power Please see below for DC power installation. The DC source is connected to the power interface on the back of the detector, the polarity of the power supply is marked in the figure on the right 3.1.4 Indicator After booting up, users can check the status indication of LED as follows.
Mars1717V Digital Flat Panel Detector User Manual 3.2 Battery Charger Installation Operation Figure Unload Battery from battery charger. Insert battery into battery charger. Note the interface position as figure. Press the battery to the bottom of battery compartment. 32 iRay Technology Co., Ltd..
Mars1717V Digital Flat Panel Detector User Manual 4. Software Setup 4.1 4.2 4.3 4.4 4.5 4.6 System requirement ............................................................................................................ 34 Environment setup .............................................................................................................. 34 Connetion Mode ..................................................................................................................
Mars1717V Digital Flat Panel Detector User Manual 4.1 System requirement iDetector is developed and deployed on Windows Operation System, it can be run on Windows XP/Windows 7/Windows 8/Windows 10, OS should install latest service pack. And requires computer memory 4 GB minimum. Firewall should be shut down to avoid commuication issue. 4.2 Environment setup Setup files and download url are included in Software Development Kit(hereafter refers as SDK) directory: Tools\env_setup. 1.
4. Software Setup 4.4 Wired Connection Setup (Service Only) Select wired network adapter that connected to the detector. Right click the network adapter. Then select properties. iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual Double click IPV4 item Default IP settings: IP address: 192.168.8.188 Subnet mask: 255.255.255.0 The IP address should be identical with Cfg_HostIP item in work_dir\Mar1717VS \config.ini file. 36 iRay Technology Co., Ltd..
4. Software Setup 4.5 Client Mode Connection To complete wireless connection configuration, user has to finish actions listed below. 1. Setup physical connection Connect one end of Gigabit Ethernet Cable to Workstation, Connect another end to LAN port of External 2. AP setup Set up the Wireless AP to: IP address 192.168.8.188 Subnet Mask 255.255.255.
Mars1717V Digital Flat Panel Detector User Manual Right click the network adapter, select properties and entered the Local connection Properties window as shown left. Double click IPV4 item Select “Obtain an IP address automatically” 38 iRay Technology Co., Ltd..
4. Software Setup Open browser and type 192.168.1.1 Log into external wireless AP Wireless setup SSID: NETGEAR_BIG_24 Security: WPA2-PSK Configure 2.
Mars1717V Digital Flat Panel Detector User Manual LAN setup Configure LAN IP address IP address: 192.168.8.1 Subnet Mask: 255.255.255.0 The AP setup is done 3. Workstation IP setup After AP setup done, the user can setup the Workstation IP address Open local network management interface 40 iRay Technology Co., Ltd..
4. Software Setup Right click the network adapter, select properties and entered the Local connection Properties window as shown left. Double click IPV4 item Set the Default IP as follows: IP address: 192.168.8.188 Subnet mask: 255.255.255.0 iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual The IP address should be identical with Cfg_HostIP item in work_dir\Mars1717VS\co nfig.ini file. 4. Panel setup Either Wired Cable or AP mode can be used to configure detector To start configuration with wired cable. It is necessary to finish 4.4, then proceed to the steps below. Connect panel to Workstation like 4.4 42 iRay Technology Co., Ltd..
4. Software Setup Click “Detector” Click “Read Config” Choose Client mode Click “Add” Type SSID and Password Click “Apply” Choose SSID and select (There will be √ occurred) Click “write config” to save parameters. iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual Turn on wireless router. Make sure there are wired connection between router and work station and IP 192.168.8.188. Click “Read WIFI Status” to check wireless transmission status, numerical value occurred means the link is up and available. Since we have chosen default SSID and password, it would connect to wireless AP immediately after powered on next time. 44 iRay Technology Co., Ltd..
4. Software Setup 4.6 iDetector software SDK supply iDetector as tool softwore: 32-bits iDetector.exe: Tools\iDetector\w32 64-bits iDetector.exe: Tools\iDetector\x64 Double click iDetector.exe to run the software. For different software version, the UI maybe have little difference. If change, forgive us for not issuing a separate notice. Users can refer to 903-341-14_iDetector_UserManual_EN_A4 for specific operation methods.
Mars1717V Digital Flat Panel Detector User Manual 4.7.1 HomePage The main function in this page is to connect detector. C B A A Item Function description Name Display the name of detector SN Display the SN of detector Product Type Display the type of detector State Display the connection state (Bind, Unknown, Ready etc.) B Button Function description Connect Click this button to connect the selected detector. Close Click this button to disconnect the selected detector.
4. Software Setup 4.7.2 Acquire Page This page is used to acquire image under different work mode, and user can choose correction mode too. When acquire image finished there will be a preview image shown on the screen.The propertities of image is displayed on the left of preview image. On the right of preview image there is a list to show thumbnail of images. User can select it and double click to see for detail. User can rotate, reverse or mirror image. User can get the value of AVG and SNR by ROI tool.
Mars1717V Digital Flat Panel Detector User Manual Correction Menu Description HW- Do hardware PostOffset correction for image if PostOffset checked (Only for Mars detector) Gain HWGain Do hardware Gain correction for image if selected Defect HWDefect Do hardware defect correction for image if checked Offset (for Mars and Mercu detector) Acqurie Button Description PrepAcq Clear and acquire Save Save image, the format is raw and tiff Poweroff shutdown detector Image Properties& Descript
4. Software Setup Please refer to 903-341-14_iDetector_UserManual_EN_A4, for specific operation guide. 4.7.4 Calibrate Page Offset, Gain, Defect calibrate files can be generated and managed in this page. Please refer to 903-341-14_iDetector_UserManual_EN_A4, for specific operation guide. Users can also refers to chapter 5.2 for panel Correlation and Calibration tutorial. 4.7.5 Local Page The idetector Software includes a local images display tool that allows users to look up local images.
Mars1717V Digital Flat Panel Detector User Manual 4.8 List of the HAZARDOUS SITUATIONS resulting from a failure of the ITNETWORK a) The operating system is not compatibility; b) Change or upgrade the software failed; c) The compatibility of the interface; d) The data transfer protocol error; e) The inconsistent of interface or format leads to data distortion; f) 50 The data output failed; iRay Technology Co., Ltd..
Mars1717V Digital Flat Panel Detector User Manual 5. Operation 5.1 MAIN OPERATION ............................................................................................................... 52 5.1.1 Steps for acquiring image ......................................................................................... 53 5.1.2 Software Mode............................................................................................................ 54 5.1.3 After Use .........................................
Mars1717V Digital Flat Panel Detector User Manual Mars1717VS provides SDK for user to integrate panel into their DR system. Additionally, it also provides an application for demonstration, i.e. iDetector. User can use iDetector to control panel without DR system. 5.1 Main Operation To Acquire X-ray image is the main operation of Mars1717VS. Most importantly, panel should build synchronization with X-ray generator.Mars1717VS is born with one way to acquire X-ray image that is Software Mode.
5. Operation 5.1.1 Steps for acquiring image Make sure the hardware is connected correctly and then power on. Once powered off, please wait at least 60s before power on again Wait until initialization is complete Connect the software choose the synchronization mode Generate HWPreOffset, Gain and Defect template after the detector reaches thermal equilibrium Acquire images in the selected mode To Acquire X-ray image is the main operation of Mars1717VS.
Mars1717V Digital Flat Panel Detector User Manual 5.1.2 Software Mode 5.1.2.1 Block Diagram Software mode is the basic way to acquire x-ray image. Please see figure below for general feature HVG X-Tube FPD Work Station Workstation is a host device installed with iDetector and SDK. Chapter 3 has described how to establish connection between panels and workstation. In software mode, workstation does not control x-ray generator. Users would decide when to shoot x-ray. 5.1.2.
5. Operation 5.1.2.3 Timing Setting To set a clear scenario for programming, see diagram below for details 1. Workstation receives “prepacquire” request, send command “clear” to panel. 2. Panel receives “clear” from workstation, starts clearing leakage of panel. Meanwhile, panel send a message to workstation “Exposure Prohibited”. 3. Panel finishes “Clear” and send a message to workstation “Exposure Enable”. 4.
Mars1717V Digital Flat Panel Detector User Manual 10. Panel completes image acquisition and begins to send data to workstation. 11. Workstation receives all image data from panel If Hardware Post offset and Hardware calibration is selected, image got would be preview image. After step11, panel would do another dark image acquisition. With both light image and dark image, panel completes all the correction and calibration process. Finally, panel uploads processed image to workstation.
5. Operation 5.1.3 After Use 1. Disconnect the software 2. Power off 3. Keep it clean 4. Store under specified conditions iRay Technology Co., Ltd.
Mars1717V Digital Flat Panel Detector User Manual 5.2 Correction and Calibration Template Generation The correction and calibration should be performed after installation and it is recommended to perform the new correction and calibration after any major change on the system settings and hardware configuration. On the other hand, it is also recommended to do the correction and calibration in each 6 months. 5.2.
5. Operation 3) After exposure, click “Acquire” to get the light image; 4) If the value meets the expected value, click “Accept”, then get the other 11 images; 5) If the value does not meet the expected value, please do not click the “Accept”, and adjust the exposure dose, then click the “PREP” to get light image again; 6) After getting 12 images, click “Generate” to generate gain template. Notes: 1 please use hardware post offset correction. 5.2.
Mars1717V Digital Flat Panel Detector User Manual Click “Local File” button in “Local File” UI, choose the specified file In this page user can open the image files saved in local, the file format can be raw, tiff, dft. When the software is disconnected to detector, the file still can be opened. Click “Load File”, there will be an open file wizard. Select file and click open or double click the file. The tiff file will be opened directly.
5. Operation Width Image width Height Image height Rotate the image clockwise, 90 degrees every time. Rotate the image anticlockwise, 90 degrees every time. Mirror Open or close mirror ROI ROI tool, to view the image of the AVG, SV, SNR and other parameters. Press "ctrl" key, can create several ROI areas. WW/WL Auto adjust WW/WL based on selected area by right button of mouse. 5.
Mars1717V Digital Flat Panel Detector User Manual The dialog box shows the version of the current firmware Click “Browse” to choose the firmware file to upgrade, the extension of the file is .ifirm After choosing the file, the lower dialog box shows the version of the new firmware, user should check the information and click “Start Upgrade” After the upgrade process is finished, power-cycle the detector 62 iRay Technology Co., Ltd..
Mars1717V Digital Flat Panel Detector User Manual 6. Regulatory Information 6.1 Medical equipment safety standards ..................................................................................... 64 6.2Guidance and manufacture’s declaration for EMC ................................................................ 66 6.2.1 EMI Compliance Table ............................................................................................... 66 6.2.2 EMS Compliance Table ....................................
6. Regulatory Information 6.
Mars1717V Digital Flat Panel Detector User Manual ANSI/AAMI ES60601- Medical electrical equipment – Part 1: General 1:2005/(R)2012+A1:2012+C1:2009/(R)20 requirements for basic safety and essential 12+A2:2010/(R)2012 performance Medical electrical equipment –Part 1: General CAN/CSA-C22.2 No.
6. Regulatory Information EN 62304:2006/AC:2008 Medical device software – Software life-cycle processes Medical devices – Application of usability engineering EN 62366:2008 to medical devices ANSI/AAMI ES60601-1:2005+ Medical Electrical Equipment – Part 1: General Amendment 1:2012+ Amendment 2:2010 requirements for safety and essential performance CAN/CSA C22.2 No.
Mars1717V Digital Flat Panel Detector User Manual Radiated RF EM IEC 61000-4-3 field 3V/m 80MHz-2.
6. Regulatory Information Electrical fast transients/burst Immunity test levels Basic EMC Phenomenon standard Professional healthcare facility environment ±2 kV IEC 61000-4-4 100kHz repetition frequency Surges IEC 61000-4-5 ±0.5 kV, ±1 kV IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV Line-to-line Surges Line-to-ground 3V, 0.15MHz-80MHz Conducted disturbances IEC 61000-4-6 induced by RF fields 6V in ISM bands between 0.15MHz and 80MHz 80%AM at 1kHz 0% UT; 0.
Mars1717V Digital Flat Panel Detector User Manual Mars1717VS conforms to this EN60601-1-2:2015 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: The use of accessories, transmitters and cables other than those specified by this User Manual, with the exception of accessories and cables sold by iRay of Mars1717VS as replacement parts for inner components, may result in increased emission or decreased immunity. 6.
6. Regulatory Information The panel may not cause harmful interference. The panel must accept any interference received, including interference that may cause undesired operation. The panel generates, uses, and radiates radio frequency energy and, if not installed and used in accordance with the instruction, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in an installation.
Mars1717V Digital Flat Panel Detector User Manual 7. Trouble Shooting TROUBLE SHOOTING ....................................................................................................... 71 iRay Technology Co., Ltd.
6. Regulatory Information Please refer to service manual. If the problem persists, turn off the panel and contact iRay service department (service@iraygroup.com). We would provide the best service. 72 iRay Technology Co., Ltd..
Mars1717V Digital Flat Panel Detector User Manual 8. Service Information 8.1 Product Lifetime .................................................................................................................... 74 8.2 Regular Inspection and Maintenance .................................................................................... 74 8.3 Repair .................................................................................................................................... 74 8.
Mars1717V Digital Flat Panel Detector User Manual 8.1 Service Office Info 8.2 Product Lifetime The estimated product lifetime is up to 5 years under appropriate regular inspection and maintenance. 8.3 Regular Inspection and Maintenance In order to ensure the safety of patients and operator, to maintain the performance and reliability of the panel, be sure to perform regular inspection at least once a year. If necessary, clean up the panel, make adjustments or replace consumables such as fuses etc.
Appendix 8.5 Replacement Parts Support Main parts (parts required to maintain the function of the product) of this product will be stocked for 5 years after discontinuance of production for repairing. iRay Technology Co., Ltd.
8. Service Information 9. Appendix Appendix A Information of Manufactures .................................................................................... 77 Appendix B Information of Medical Device Directive European Representative ........................ 77 76 iRay Technology Co., Ltd..
Mars1717V Digital Flat Panel Detector User Manual 9.1 Appendix A Information of Manufactures COMPANY: iRay Technology Taicang Ltd. ADDRESS: No.33 Xinggang Road, Taicang Port Economic and Technological Development Zone, Jiangsu, China ZIP CODE: 215434 TELEPHONE: +86 0512-53690872 FAX: +86 0512-53690872 HOMEPAGE: WWW.IRAYGROUP.COM 9.
8. Service Information CC Regulations: Contains module’s FCC ID:2ACHK-01070189 This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
Appendix This is because the device is designed to operate at multiple power levels so as to use only the power required to reach the network. In general, the closer you are to a wireless Base station antenna, the lower the power output. The exposure standard for wireless devices employing a unit of measurement is known as the Specific Absorption Rate, or SAR. The SAR limit recommended by the ICNIRP used by the general public is 2.
8. Service Information be installed and operated with minimum distance of 0 cm between the radiator and your body. This device and its antenna(s) must not be co-located or operating in conjunction with any other antenna or transmitter. Cet appareil est conforme aux Normes RSS d'Industy Canada.
Appendix 0.106 W/kg. Bien qu'il puisse exister des différences entre les niveaux de SAR selon les dispositifs et les emplacements où ils sont utilisés, tous répondent aux exigences Gouvernementales. La valeur SAR déclarée conforme est une distance de 0 mm entre l'unité et le corps humain. Eloignez cet appareil à une distance d'au moins 0 mm de votre corps pour vous assurer que le niveau d'exposition aux RF est conforme ou inférieur au niveau indiqué.
8. Service Information End of document 82 iRay Technology Co., Ltd.