User's Manual Part 3

8-15
Problem
Possible Cause
Solution
Printer out of paper or
paper jam
1) Review section 7 of
the operators
manual
8.6.4. Troubleshooting MRI Related Problems
Problem
Possible Cause
Solution
1
Artifact on MR images.
Loose connection to cord
when running on AC
power.
Tighten AC cord connection
to Monitor and power
supply.
Faulty Hardware.
1) Switch to battery
power.
2) Disconnect the power
cable from the 3880
monitor.
3) Disconnect AC power
cord from the AC
outlet in the MRI room.
4) Inspect components
damage.
5) Remove all power
cables from the room.
6) If artifact disappears
then replace faulty
hardware.
7) If artifact is still present
repeat steps for the
3886.
8) If artifact still appears
contact IRadimed
technical service.
8.6.5. Troubleshooting Vital Sign Related Problems
Problem
Possible Cause
Solution
1
ECG Vital Signs are not
performing as expected
1) Ensure the ePOD is
on the same
wireless network
channel as the 3880.
2) Ensure the ECG
parameter is
enabled.
3) Inspect all ECG
components for
damage and replace
as necessary.
4) Review section 6.1
of the operators
manual
5) Contact your local
representative to
schedule
applications training.