User's Manual Part 3

8-2
8.1.2. Checking a Battery
The user replaceable battery packs utilized in the 3880 patient monitor can be checked outside
of the monitor. This feature allows facilities with multiple batteries to proactively select a battery
that has a desired charge level prior to inserting it into the system. To check a batteries charge
follows these steps:
1. Locate the button on the rear corner of the battery (1)
2. Depress the button and the LED indicators will momentarily show the charge status. (2)
8.2. Performance Checks
8.2.1. Daily in between tasks
Check that system components and accessories do not have any visual defects such as
cracks or loose parts.
Check that the system components as described in the Care and Cleaning 8.4 part of
this section were cleaned properly after the previous use.
All supplies and accessories are not passed their expired date.
Check that batteries are properly inserted and charged.
Ensure all system components are wirelessly communicated
Wipe the monitor, PODs and accessory surfaces
Change all airway patient accessories.
Ensure that you are using correct accessory sizes for each patient and that they are
properly connected.
Check that all accessories, cables and monitor parts are clean and working properly.
Check that you have selected desired parameters to be displayed in digit and waveform
fields.
Check that the trends of the previous patient are erased.
Monitor patient type mode and alarm limits are suitable for the patient.
Check the compatibility of the 3880 monitor, sensors and cables before use
8.2.2. Regular Inspection and Verification
Wireless Communication
Place the PODs and 3880 Monitor inside the MRI room suite with the door shut. Place
the 3885-T Remote Tablet in the control room in its desired location and check that the
communication performance is acceptable.
ECG
Check that the message 'Leads Off' disappears and the waveforms are displayed when
the lead wires are connected to the patient.
Pulse Oximetry
Check that the Red Light is visible in the sensor
Check that the SpO
2
value is displayed and the message “Probe Off” disappears when
the sensor is connected to the patient
CO
2
(CO2 only or 3886 Multi-Gas system)