User's Manual Part 2
2-54
16. Messages and Alerts area
NOTE
• If the touchscreen is not responsive or blank, remove from use and refer the monitor to
qualified service personnel.
2.2.11.5 Using Control Switch Hard Keys
1. SETTINGS: Access monitor setup menus
2. ALARM OFF - Standby button: Indefinitely pauses all alarms and terminates
automatic NIBP measurements
3. TRENDS: Trend Screen access and adjustment
4. PRINT: Prints to optional recorder in the Base Station
5. NIBP START/STOP: Initiates a NIBP measurement when one is not in progress, or
stops an NIBP in progress. Holding START/STOP button for 3 seconds initiates STAT
readings.
6. AUDIO ALARM OFF – Alarm Silence button: multi-function audible alarm control,
resets sounding of alarm, pauses for 120 seconds the alarm sound, or re-enables alarm
sound capability.
2.2.11.6 Using Menus (Touch Screen controls)
1. Menu Title: Description correlating the menu choices
2. SELECTION BUTTON: Touching will make a desired selection or open up a new menu
3. DROP DOWN BUTTON 1: Touching will enable a pop up menu to make a selection
from multiple choices
4. DROP DOWN BUTTON 2: Touching will enable a pop up menu to make a selection
from multiple choices
5. BACK BUTTON: Closes menu and returns user to previous menu or main monitoring
screen
6. TOGGLE SWITCH: Touching will toggle the desired selection between Enabled and
Disabled
1
2
3
4
5
6
1
2
3
4
5
6