User's Manual Part 2

2-51
2.2.10.5. NIBP Lines
The NIBP feature utilizes the oscillometric method for measuring and displays systolic, diastolic,
and mean arterial pressures, and pulse rate.
To connect the NIBP Line:
1. Locate the NIBP line (1) and the NIBP receptacle on the 3880 monitor (2)
2. Position the NIBP line with the NIBP receptacle so they align
3. Firmly insert the NIBP line onto the corresponding receptacle ( 2) until the NIBP line is
secure, notice a snap when locked into place.
4. Ensure that NIBP line is secure into the monitor and not loose
! CAUTION
Use only NIBP accessories specifically designed and authorized for use with the
IRadimed 3880 system. Refer to section 9.2 for a complete list of available accessories.
2.2.10.6. Temperature Fiber Optic Cable
The 3880 system can be configured with an optional temperature channel to continuously
measure either a patient’s surface or body temperature.
To connect the Temperature Cable:
1. Locate the temperature cable (1) and the temperature receptacle on the 3880 monitor
(2)
2. Position the temperature cable with the temperature receptacle (2) so they align
3. Gently insert the temperature cable into the corresponding receptacle and rotate the
outer locking ring clockwise until it stops.
4. Ensure that temperature cable is secure into the monitor and not loose.
! CAUTION
The temperature sensors are constructed of fiber-optic glass and must always be
handled with care to prevent damage. Improper handling can result in inaccurate
readings.
1
2
1
2
1