User's Manual Part 2
2-62
NOTE
• Wireless Channel cannot be stored under a user setting. The 3880 system components
will power on with the last channel that was used. See 2.2.12.9 Auto settings Memory.
• Settings modified during use will be retained and used for short power on to off to on
cycle of < 1 minute.
2.2.12.7.1 Saving a New Setting
To save a custom user setting, prepare the desired setup on the 3880 monitor or 3885-T
Remote Tablet before entering the Edit User Settings menu. The first user setting (A) is
considered the default setting and will automatically load upon system power on. User settings
can be saved to a USB drive to ease the standardization of settings.
To create and store a user setting follow these steps:
1. Configure the system, parameters, recorder and alarms for the stored setting
2. Press “SETTINGS” button
3. Select “Store / Recall Setups“
4. Select the desired position to store
• Note: the first user setting (A) is the default setting and will automatically load upon
system power on
5. Touch the “Store” button to save the settings to that position
6. Touch the Back button to close the menu
NOTE
• The system will not store alarm Limits or sounds to “OFF”
• The settings A, B, C, D and E are all set to Factory Defaults upon initial shipment from
IRadimed.
2.2.12.8. Service Setup
The service setup menu contains technical utilities to test the system’s performance. Please
contact your service representative or reference the 3880 service manual for details.
2.2.12.9.
Auto Settings Memory
(The following settings are automatically stored as set and recalled at power up)
• Last communication channel
• Notch Filter
• Language