User's Manual Part 3
6-18
6.3.3. CO2 and Multi-Gas Patient Application
To apply the CO
2
/ Agent accessories follow these steps:
1. Ensure hardware is fully warmed up
2. Connect exhaust port of the monitor to the scavenge system using part number 1846
3. Connect an unused
IRadimed sampling line to the 3880 CO2 gas inlet port
4. Connect the sampling line to either the airway adapter or in the case of nasal cannula, to
the in 6.3.3.1
5. Review the values for consistency with patient
! CAUTION
• In the patient’s sampling line can contain infectious elements, and can be a potential
hazard for the 3880 monitor user(s). Use caution when disconnecting and disposing of
used sampling lines.
• Always use the supplied microbial filter to keep biological and water condensate
contamination out of the monitor.
• Do not damage, separate or occlude sampling line
6.3.3.1. Intubated Patients, with use of airway adapter, suitable for all patients, with
or without anesthetic agents
6.3.3.2. Cannula Patients, use for CO2 measurement only, no agents
NOTE
Select size A, P, or I for best mechanical fit to patient nasal size.
P/N 1849 Sample line,
Nomoline, mates only with
P/N 3886 CO2/Agents unit
To ventilator
Airway Adapter
P/N 1842 A, P, or I
Sample line