User's Manual Part 3

6-14
Adult SpO
2
Off, 50-99
90%
Off
70-99, Off
Pediatric SpO
2
Off, 50-99
90%
Off
70-99, Off
Neonatal SpO
2
Off, 50-99
85%
99%
70-99, Off
Adult Pulse Rate
Off, 30-239
50
120
50-240, Off
Pediatric Pulse Rate
Off, 30-239
75
180
50-240, Off
Neonatal Pulse Rate
Off, 30-239
90
200
50-240, Off
6.2.6. SpO
2
Messaging
Message
Trigger Condition
SpO2 Inop
SpO
2
hardware or software failure detected
SpO2 Searching
Searching for patient pulse
SpO2 Probe Off
Sensor is not properly attached to the patient
SpO2 Low Batt
15% of battery life is remaining in the oPOD
SpO2 Bad Probe
SpO2 sensor not compatible or damaged
SpO2 No Probe
SpO2 sensor is disconnected from the oPOD
Low Perfusion
Low perfusion detected