User's Manual Part 3
6-8
• See the WARNING at 6.1.3 regarding the non-standard lead placement and the
foreseeable changes the ECG waveform due to placement.
6.1.4.3. HR Source
HR Source permits the user to select the vital sign source that will to be used to produce the
heart rate displayed in the ECG vital sign box. The following options are available:
Option
Corresponding Vital Sign
ECG
Electrocardiogram
SpO
2
Pulse Oximetry
NIBP
Most recent NIBP Measurement
6.1.4.4. Gating Source
The HR Source will not affect the gating signal that the 3880 monitor outputs.
• ECG – The gating pulse will be produced using data from the ECG vital sign, lead Trace
A only.
• SpO
2
– No gating pulse will be produced using data from the SpO
2
vital sign.
• NIBP – No gating pulse will be output from the 3880
6.1.4.5. Gating Cable
The gating cable is a physical interface between the MRI scanner and the 3880 monitor. To
interface with the scanner follow these steps:
1. Connect the gating cable to the 3880 gating port
2. Locate the ECG gating cable that came with the scanner and instead of connecting it to
the patient electrodes connect it to the IRadimed 3880 gating cable. Match the
corresponding lead colors of the scanner gate cable to those of the IRadimed gating
connections.
NOTE
• The scanners ECG gating cable will always be used with this cable.