User's Manual Part 3
2-61
2.2.12.7. Store/Recall of User Setup
This feature allows you to store multiple user setups and to select one for a default power up
setting. Storage of different procedures, patient types and users are available by customizing
the following:
• Alarms
1. Minimum and Maximum Limit Values, not including
‘OFF’
2. Latched or unlatched
3. Alarm volume level, not including OFF or 0 sound
• Patient Type
1. Adult
2. Pediatric
3. Neonatal
• SpO
2
1. SpO
2
parameter on or off
2. Sweep Rate
• ECG
1. ECG Lead 1 on or off, Lead 2 OFF
2. Selected lead
3. Scale setting
4. Sweep Rate
5. Filter mode
6. QRS tone volume
7. Heart rate source
• NIBP
1. NIBP parameter on or off
2. Automatic time interval
• Temperature
1. Temperature parameter on or off
2. Unit
• CO
2
1. CO
2
and Respiration on or off
2. Scale
3. Unit
4. Resp Source
5. Sweep Rate
• Trends
1. Interval
• Recorder
1. Selected waveform(s)
2. Trace Delay
3. Speed
4. Run Time
5. Auto Strip
• Display Brightness
• Dynamic Trend
1. ON / OFF
2. Ave. Period
• MAC
FACTORY DEFAULT
Unlatched
100 %
Adult
ON
25
ON (Lead 1 only)
II
10
25
MRI
50 %
ECG
ON
Off / Manual
OFF
C°
OFF
40
mmHg
CO2
6.25
ON
3 min
ECG
8 sec
25
30 sec
OFF
75 %
OFF
5 min
OFF