User's Manual Part 3
2-59
When connection is established, the channel link signal indicator “bars” illuminate green in the
top left channel icon display area
2.2.12.4.5 Adjusting the 3885-B Base Station Wireless Channel
1. Complete the steps in section 2.2.12.4.4 to assign the channel on the 3885-T Remote
Tablet
2. Dock the 3885-T Remote Tablet to the 3885-B Base Station and observe the LED
display on the base station to confirm connection – OR -
3. Press the “Channel Select” button on the 3885-B Base Station until the channel LED
shows the desired channel number.
4. Confirm all 3880 system components are now communicating by observing the
simultaneous flashing (4 seconds) of the green LED at both the 3885-T Remote Tablet
and 3880 Monitor
2.2.12.5. Recorder Setup
A 3880 system communicating with the optional 3885-B Base Station strip chart recorder can
provide hard copies of up to two waveforms, trend information and patient data reports.
2.2.12.5.1 Recorder Settings
The optional recorder can be adjusted to suit a variety of needs. It can be configured to print up
to 2 waveforms simultaneously as well as operation features such as the trace delay, sweep
speed and run time.
• The printable waveforms are: ECG trace A, ECG trace B, SpO
2
, CO
2
• The trace delay adds a buffer to account for a delayed reaction when pressing the print
button. The 3880 trace delay can be configured for real time as well as 4, 8 and 16
second delays.
• The system can be configured to automatically print during an alarm condition.
• The recorder can be configured to print 8, 12, 16, 20 and 30 second strips once the print
button is pressed.
• The recorder sweep speed can be adjusted to 25 or 50 mm/s