User's Manual Part 2
10-26
Note 1: Negligible interference, effect included in the specification “Accuracy, all conditions” above.
Note 2: Negligible interference with N
2
O / O
2
concentrations correctly set, effect included in the
specification “Accuracy, all conditions” above.
Note 3: Interference at indicated gas level. For example, 50 vol% Helium typically decreases the CO
2
readings by 6%. This means that if measuring on a mixture containing 5.0 vol% CO
2
and 50
vol% Helium, the actual measured CO
2
concentration will typically be (1-0.06) * 5.0 vol% = 4.7
vol% CO
2
.
Note 4: According to the EN ISO 80601-2-55:2011 standard.
Note 5: In addition to the EN ISO 80601-2-55:2011 standard.
Note 6: CO2 Only internal option is contra-indicated for use with anesthetic agents, no interfering gas
testing or specification per EN ISO 80601-2-55 standard.